GWInstek GLC-9000 Safety Testers

Brief description

The GLC-9000, leakage current tester, is used to perform leakage current ( or called touch current) tests on medical electrical (IEC 60601-1) and general purpose electric & electronic (IEC 60990) equipment. This tester engages with nine measurement networks ( or called Measuring Device) to provide the simulation of human body whilst the EUT ( equipment under test) is taking a leakage current testing, in compliance with the specific standards or regulations such as IEC, UL, JIS…etc..

Properties

  • Touch Panel with Color LCD Display
  • 9 Different Measurement Network to Simulate
  • the Resistance of Human Body
  • 8 Different Types of Leakage current for Medical
  • & General Electrical Equipment
  • Various Leakage Current Measuring Mode:
  • DC/AC/AC+DC/AC Peak
  • Upper & Lower Limitation for PASS/FAIL Judgment
  • Meter Mode with SELV/CONV Function
  • 50 Preset Test Conditions Conform to IEC 60990;
  • 30 Sets Memories for Customer Defined
  • Various Standard Interfaces:
  • RS232/GPIB/USB Host & Device/EXT I/O

Technical attributes:

Metering Mode AC/DC/AC+DC; AC peak
Voltage 85V ~ 300V
Current 0.5A ~ 10A
Interface USB host; RS-232C/GPIB/EXT I/O/USB device
Power supply range AC 100V/120V/220V/230V±10%,50/60Hz
Dimensions (mm) 330 (W) x 150 (H) x 350 (D) mm
Volume 5000 g

PRODUCT INFORMATION

Product Code: 350301
Manufacturer:
GW Instek

Availability

Orderable

PRICE

Net price: 820 144 Ft
Gross price (27% VAT): 1 041 583 Ft
Request an Offer

Add to basket

Pieces/Packages: 1 pcs
Minimum order quantity: 1 pcs